Baxter 76301 - PUMP, INFUSION, BAXTER, FLOGARD 6301, DUAL, EACH

Flo-Gard 6301, Dual Channel Volumetric Infusion Pump

Principal Features

The Flo-Gard 6301 Dual Channel Volumetric Infusion Pump is an electromechanical device used for the intravenous infusion of liquids at user-selected rates. The device contains two linear peristaltic pump heads, allowing it to simultaneously infuse two different solutions. Each pump head is independently programmable and permits infusion of primary and secondary medication programs. The secondary program automatically switches over to the primary program when secondary infusion is complete (automatic piggybacking).

The device operates on standard 115 VAC 60 Hz electrical power, or on its self-contained rechargeable battery. It is portable and has a panel lock-out feature to prevent patient tampering. It is designed for use with Baxters standard administration sets which contain an s as the last character of the code number, for example, 2C5537s. When infusing solutions through a central line catheter, sets with Luer lock adapter should be used. Sets with a Flashball@ device are not recommended in these applications.

The primary rate of infusion is selectable from 1 to 1,999 mL in 1 mL/hr incrementsand 1 .O to 99.9 mL/hr in 0.1 mL/hr increments. The secondary rate is selectable from 1 to 999 mL in 1 mL/hr increments and 1 .O to 99.9 mL/hr in 0.1 mL/hr increments. The volume to be infused (VTBI) is also selectable from 1 to 9,999 mL in 1 mL increments and 1 .O to 99.9 mL in 0.1 mL increments.

The total volumes infused from primary and secondary programs are added together and accumulated and can be displayed on demand. The primary and secondary VTBIs are independently decremented and displayed. Upon completion of the primary VTBI, the pump automatically switches to a keep vein open (KVO) rate. If the pump is started on a secondary rate and VTBI, the pump will change to the primary rate when the secondary program is completed. Either pump may be stopped at any time by depressing the STOP key unless the device is in the lockout mode.

Technical Specifications

ItemCharacteristic
DescriptionDual channel linear peristaltic volumetric infusion pump
Administration SetBaxters standard administration set with "s" suffix
Keep Vein Open (KVO) rate5 mL/hr or programmed rate, whichever is less
Battery12 Volt, 3.2 Ah sealed lead acid
Battery Life
  • Approximately 6 hours with one pump running at a rate from 1 to
    1400 mUhr using a fully charged battery
  • Approximately 4 hours with both pumps running at a rate from 1 to
    1400 mL/hr using a fully charged battery
Battery Recharge8 hours for 80% recharge
AC Power Requirements11 O/f20V, 60 Hz
Power Cord2.75 m (9.0 ft.) long, with Hospital Grade plug
Leakage CurrentTypically less than 50 microamps (using UL-544 specified test methods)
Power Consumption50 watts
Weight8.2 kg (18 Ibs)
Flow Rate Range
  • Primary program: 1 - 1,999 mL/hr in 1 mL/hr increments and 1 .O - 99.9 mUhr in 0.1 mUhr increments. Upper limit can be reduced by authorized service personnel.
  • Secondary program: 1 - 999 mUhr in 1 mUhr increments and 1 .O - 99.9 mUhr in 0.1 mUhr increments.
VT81 RangeI - 9,999 mL in 1 mL increments and 1 .O - 99.9 in 0.1 mL increments for both primary and secondary. Upper limit can be reduced by authorized service personnel.
Air-in-Line DetectionFactory set to NORM which causes the device to alarm on air bubbles approximately 75pL or larger. The MIN setting causes the device to alarm on bubbles approximately 50 PL or larger.
Fuse0.5 A Slo-610
Nurse Callg-pin D connector, Pin 1: N/C (normally closed), Pin 4: N/O (normally open), Pin 9: Common Contact rating: 0.4A at 30 VAC, 2A at 30 VDC resistive load (internal connection required)