Fisher & Paykel OPT316 - CANNULA, OPTIFLOW, JUNIOR, INFANT, 20 PER/CS

Part Number Fisher & Paykel OPT316
SKU Number CIA7001709
Sell Unit CASE
Ships Within 2-3 Days
List Price $3,135.03
Enter a valid quantity
Product Description

Fisher & Paykel OPT316 - CANNULA, OPTIFLOW, JUNIOR, INFANT, 20 PER/CS

Optiflow Junior Nasal Cannula - Infant (Flow up to 20L/min), 20/box

F&P Optiflow Junior provides a revolutionary step between low-flow oxygen therapy and CPAP. In some clinical scenarios it has even been found to reduce the requirement for CPAP and intubation. By combining an anatomically contoured nasal cannula with comfortable humidified flow, your neonatal and pediatric patients can receive vital oxygen and flow support thats easier for clinicians, effective for their patients and kinder on parents.

  • A cannula range designed specifically for the delicate anatomical features and flow requirements of neonates, infants and children
  • Award-winning design
  • Breathable FlexiTube minimizes troublesome mobile condensate
  • Kink-proof and crush-resistant FlexiTube makes therapy delivery safer
  • Adhesive Wigglepads easily maintain cannulae stability on cheeks
  • Easy-click connector rotates for convenient movement of patient and breathing circuit
  • Color coded for easy size identification

Nasal cannula OPT31X

Components & Composition

Regulatory

Superior Infant Care

Precious new life deserves the best possible start. From the first breath the F&P Infant Respiratory Care Continuum facilitates the evolution from immature and supported lung function to respiratory independence. At every point of the Care Continuum, humidified solutions help to emulate the natural physiological balance in healthy, mature lungs. As an infants needs change, so does the configuration of the therapy system. As a result caregivers can nurture life, confident they are using the best therapy solutions, delivered in the most efficient ways using F&P Optiflow Junior.

Application & Removal

To Apply Cannula

  • Connect system to gas source. Place hand close to the nasal prongs to ensure that there is airflow exiting the prongs.
  • Remove adhesive backing and place the nasal cannula in the nares ensuring that there is at least a 2 mm (0.08 inches) gap from the base of the prongs to the septum.
  • Ensure the cannula size is correct. Prongs should not fill the nares and a clear gap should be visible around each prong.

Checks During Operation

  • Check patients septal integrity.
  • Ensure that the tubing is not applying excessive pressure to the ears and face.
  • Ensure that there are no secretions in the nares that could cause the prongs to seal in the nares.
  • Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient. Ensure that all connections are secure during use.

System Specifications

  • MR850 Humidifier with RT330 delivery tubing and chamber kit. MR850 Humidifier in invasive mode. OR AIRVO 2/myAIRVO 2 Humidifier with 900PT531 delivery tubing and chamber kit.
  • Ambient Range: 18 26 oC.
  • This product does not contain natural rubber Latex, PVC or Phthalates (DEHP, DBP, BBP).

This product is intended to be used for a maximum of 7 days.

Contraindications

This therapy must not be used where continuous positive airway pressure (CPAP) is contraindicated.

Warnings

  • Patient monitoring is recommended.
  • Tubing contains stainless steel. Do not use in Magnetic Resonance environment.
  • Tubing poses a potential strangulation hazard.
  • Do not use if packaging is not sealed.
  • MR850: This device must be used with the pressure relief valve supplied with the RT330 Optiflow Tubing Kit. Failure to use this safety device can allow unlimited pressure to be delivered to the patient.
  • Do not stretch or crush tube.
  • Do not soak, wash, sterilise, or re-use this product.
  • Reuse may result in transmission of infectious substances, interruption to treatment, serious harm or death.
  • Use of a non-approved accessory could impair performance or compromise safety.
  • Do not use near a naked flame.
  • Check for condensate regularly. Drain as required.